Data Foundry

Medical device recalls 2013

APLIF Implants and Instruments — Class II medical device recall Z-1157-2013

Terminated recall by Spinal Solutions, LLC, reported 2013-05-01, distributed nationwide. Spinal Solutions is recalling the APLIF system because it is not supported by adequate testing and documentati

Recall numberZ-1157-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSpinal Solutions, LLC, Murrieta, CA, United States
Recall initiated2013-03-08
Classified2013-04-22
Reported by FDA2013-05-01
Terminated2014-03-26
DistributedNationwide (US)
Quantity220 units
Reason classesdevice malfunction
Model numbers184008-10

Product

APLIF Implants and Instruments, Part Numbers: 184008-10 Anterior Psoas Lumbar 18 x 40 x 08mm - 10degree 184010-10 Anterior Psoas Lumbar 18 x 40 x 10mm - 10degree 184012-10 Anterior Psoas Lumbar 18 x 40 x 12mm - 10degree 184014-10 Anterior Psoas Lumbar 18 x 40 x 14mm - 10degree 184016-10 Anterior Psoas Lumbar 18 x 40 x 16mm - 10degree 184008-00 Anterior Psoas Lumbar 18 x 40 x 08mm - 00degree 184010-00 Anterior Psoas Lumbar 18 x 40 x 10mm - 00degree 184012-00 Anterior Psoas Lumbar 18 x 40 x 12mm - 00degree 184014-00 Anterior Psoas Lumbar 18 x 40 x 14mm - 00degree 184016-00 Anterior Psoas Lumbar 18 x 40 x 16mm - 00degree 184510-08 Anterior Psoas Lumbar 18 x 45 x 08mm - 10degree 184510-10 Anterior Psoas Lumbar 18 x 45 x 10mm - 10degree 184512-10 Anterior Psoas Lumbar 18 x 45 x 12mm - 10degree 184514-10 Anterior Psoas Lumbar 18 x 45 x 14mm - 10degree 184516-10 Anterior Psoas Lumbar 18 x 45 x 16mm - 10degree 184510-00 Anterior Psoas Lumbar 18 x 45 x 10mm - 00degree 184512-00 Anterior Psoas Lumbar 18 x 45 x 12mm - 00degree 184514-00 Anterior Psoas Lumbar 18 x 45 x 14mm - 00degree 184516-00 Anterior Psoas Lumbar 18 x 45 x 16mm - 00degree 185008-10 Anterior Psoas Lumbar 18 x 50 x 08mm - 10degree 185010-10 Anterior Psoas Lumbar 18 x 50 x 10mm - 10degree 185012-10 Anterior Psoas Lumbar 18 x 50 x 12mm - 10degree 185014-10 Anterior Psoas Lumbar 18 x 50 x 14mm - 10degree 185016-10 Anterior Psoas Lumbar 18 x 50 x 16mm - 10degree 185008-00 Anterior Psoas Lumbar 18 x 50 x 08mm - 00degree 185010-00 Anterior Psoas Lumbar 18 x 50 x 10mm - 00degree 185012-00 Anterior Psoas Lumbar 18 x 50 x 12mm - 00degree 185014-00 Anterior Psoas Lumbar 18 x 50 x 14mm - 00degree 185016-00 Anterior Psoas Lumbar 18 x 50 x 16mm - 00degree 185508-10 Anterior Psoas Lumbar 18 x 55 x 08mm - 10degree 185510-10 Anterior Psoas Lumbar 18 x 55 x 10mm - 10degree 185512-10 Anterior Psoas Lumbar 18 x 55 x 12mm - 10degree 185514-10 Anterior Psoas Lumbar 18 x 55 x 14mm - 10degree 185516-10 Anterior Psoas Lumbar 18 x 55 x 16mm - 10degree 185508-00 Anterior Psoas Lumbar 18 x 55 x 08mm - 00degree 185510-00 Anterior Psoas Lumbar 18 x 55 x 10mm - 00degree 185512-00 Anterior Psoas Lumbar 18 x 55 x 12mm - 00degree 185514-00 Anterior Psoas Lumbar 18 x 55 x 14mm - 00degree 185516-00 Anterior Psoas Lumbar 18 x 55 x 16mm - 00degree The Eminent Spine Interbody Fusion System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment.

Reason for recall

Spinal Solutions is recalling the APLIF system because it is not supported by adequate testing and documentation to demonstrate that it meets performance or safety standards. These inadequacies might result in product performance failures that could cause patient harm due to implant breakage, movement, or inadequate sterilization.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1157-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.