Data Foundry

Medical device recalls 2024

Medline Contro-Bulb Syringe — Class II medical device recall Z-1157-2024

Ongoing recall by MEDLINE INDUSTRIES, LP - Northfield, reported 2024-09-25, distributed nationwide. Packaging material was updated for the Bulb Irrigation Syringe (DYND20125), and additional testing done at Med

Recall numberZ-1157-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMEDLINE INDUSTRIES, LP - Northfield, Northfield, IL, United States
Recall initiated2024-01-04
Classified2024-02-16
Reported by FDA2024-09-25
DistributedNationwide (US)
Quantity4,831 units
Reason classessterility
UPC / GTIN / UDI-DI10080196031764, 40080196031765
Lot numbers96922110001, 96923100001
Model numbersDYND20125

Product

Medline Contro-Bulb Syringe, REF DYND20125; bulb irrigation syringe

Reason for recall

Packaging material was updated for the Bulb Irrigation Syringe (DYND20125), and additional testing done at Medline Corporate indicated the potential for the sterile barrier to be breached during transportation. This only affects two lot numbers (96922110001 and 96923100001).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1157-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.