Data Foundry

Medical device recalls 2016

Product 60 consists of all product under product code: HSB and same… — Class II medical device recall Z-1158-2016

Terminated recall by Zimmer Manufacturing B.V., reported 2016-03-23, distributed nationwide. As a result of the insufficient sealer calibration data for product packaged in firm's Building II between Aug

Recall numberZ-1158-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmZimmer Manufacturing B.V., Mercedita, PR, United States
Recall initiated2016-01-11
Classified2016-03-12
Reported by FDA2016-03-23
Terminated2017-04-13
DistributedNationwide (US)
Quantity1,394
Reason classessterility
Lot numbers61597004, 61597004N, 61621138, 61763506, 61827280, 61835479, 61878730, 61878732, 61878733, 61878734, 61878735, 61878736, 61893774, 61895931, 61895932, 61895933, 61895935, 61906259, 61912972, 61912975, 61912976, 61912977, 61920741, 61929350, 61929351, 61940068, 61940069, 61940070, 61940071, 61940072, 61940076, 61964034, 61968082, 61983672, 62086349, 62086350, 62086351, 62086353, 62086355, 62086356 and 21 more
Model numbers225207055

Product

Product 60 consists of all product under product code: HSB and same usage: Item no: 225207055 RECON SCREW LG 5.5MM DIA 225207555 RECON SCREW LG 5.5MM DIA 225208055 RECON SCREW LG 5.5MM DIA 225208555 RECON SCREW LG 5.5MM DIA 225209055 RECON SCREW LG 5.5MM DIA 225209555 RECON SCREW LG 5.5MM DIA 225210055 RECON SCREW LG 5.5MM DIA 225210555 RECON SCREW LG 5.5MM DIA 225211555 RECON SCREW LG 5.5MM DIA 225212055 RECON SCREW LG 5.5MM DIA 225213055 RECON SCREW LG 5.5MM DIA Usage: An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Reason for recall

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1158-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.