*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 20mm x… — Class I medical device recall Z-1163-2026

Ongoing recall by Boston Scientific Corporation, reported 2026-02-11, distributed nationwide. Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable res

Recall numberZ-1163-2026
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific Corporation, Marlborough, MA, United States
Recall initiated2025-12-19
Classified2026-02-05
Reported by FDA2026-02-11
DistributedNationwide (US)
Countries outside the USAE, AR, AT, BE, BR, CA, CH, CL, CO, CR, CZ, DE, DK, DZ, ES, FI, FR, GB, GR, HR, HU, ID, IE, IL, IN, IR, IT, JO, JP, KW, LB, LU, MA, MX, MY, NL, NO, OM, PA, PE, PK, PL, PT, PY, RO, RS, SA, SE, SI, SK, ZA
Quantity588 units
UPC / GTIN / UDI-DI08714729951100
Lot numbers35793920, 35952745, 35961004, 35995006, 35995007, 35995165, 36001950, 36001951, 36001952, 36001953, 36002902, 36002903, 36002904, 36003946, 36003947, 36003948, 36012845, 36012846, 36012847, 36012848, 36015761, 36021049, 36021050, 36021051, 36021052, 36027054, 36027055, 36027056, 36027057, 36027058, 36027059, 36027060, 36027061, 36038410, 36038411, 36040000, 36040001, 36040002, 36040003, 36040004 and 460 more

Product

*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 20mm x 10mm; Material Number (UPN): M00553560;

Reason for recall

Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.