NeuViz 16 Multi-Slice CT Scanner System — Class II medical device recall Z-1165-2013

Terminated recall by Philips And Neusoft Medical Systems Co., Ltd., reported 2013-05-01, distributed nationwide. User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was

Recall numberZ-1165-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips And Neusoft Medical Systems Co., Ltd., Shenyang, Liaoning, China
Recall initiated2013-03-04
Classified2013-04-23
Reported by FDA2013-05-01
Terminated2013-11-22
DistributedNationwide (US)
Quantity23 units
Serial numbersN16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120020, N16E120022, N16E120023, N16E120024, N16E120052, N16E120053, N16E130002
Model numbers989605858501

Product

NeuViz 16 Multi-Slice CT Scanner System, Part number (PN): 989605858501, Model: NeuViz 16. This is a computed X-ray, Tomography system. The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients

Reason for recall

User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control panel.