Toco M2734A — Class II medical device recall Z-1165-2020
Ongoing recall by Pacific Medical Group Inc., reported 2020-02-19, distributed nationwide. Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that
| Recall number | Z-1165-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Pacific Medical Group Inc., San Clemente, CA, United States |
| Recall initiated | 2019-09-11 |
| Classified | 2020-02-12 |
| Reported by FDA | 2020-02-19 |
| Distributed | Nationwide (US) |
| Quantity | 466 |
Product
Toco M2734A, that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips Avalon Smart M2734A/2735A Toco Gauge Flex, NFPHA9341
Reason for recall
Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.