Data Foundry

Medical device recalls 2020

Ultrasound Transducer — Class II medical device recall Z-1167-2020

Ongoing recall by Pacific Medical Group Inc., reported 2020-02-19, distributed nationwide. Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that

Recall numberZ-1167-2020
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPacific Medical Group Inc., San Clemente, CA, United States
Recall initiated2019-09-11
Classified2020-02-12
Reported by FDA2020-02-19
DistributedNationwide (US)
Quantity5,897

Product

Ultrasound Transducer (M2736A), that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips M2736A Avalon Smart Ultrasound Bottom Case, NFPHA9515

Reason for recall

Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.