Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac… — Class I medical device recall Z-1171-2013
Terminated recall by Newport Medical Instruments Inc, reported 2013-05-08, distributed in AL, AR, AZ, CA, CO, CT, DE, FL…. Newport Medical Instruments is conducting a voluntary recall on certain Newport Medical Instruments HT70 and H
| Recall number | Z-1171-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Newport Medical Instruments Inc, Costa Mesa, CA, United States |
| Recall initiated | 2013-04-05 |
| Classified | 2013-04-26 |
| Reported by FDA | 2013-05-08 |
| Terminated | 2015-03-10 |
| Distributed | AL, AR, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NH, NJ, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, WI |
| Countries outside the US | AE, AR, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CR, EC, EG, FR, GB, GR, HK, HR, HU, ID, IE, IL, IN, IT, JP, KR, LB, LK, LT, MX, MY, NO, NP, OM, PE, PH, PL, PY, QA, SA, SG, TH, TR, TW, VE, VN, ZA |
| Model numbers | HT70 |
Product
Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries Model number: HT70 and HT70 Plus. The HT70 and HT70 Plus ventilator systems are intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces.
Reason for recall
Newport Medical Instruments is conducting a voluntary recall on certain Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries due to customer reports of Newport HT70 and HT70 Plus ventilators alarming and going to internal backup battery sooner than expected while the ventilator is being operated on Power Pac battery.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1171-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.