Maquet Getinge Group battery modules for use with Maquet SERVO-i… — Class I medical device recall Z-1177-2013
Terminated recall by Maquet Cardiovascular Us Sales, Llc, reported 2013-05-08, distributed nationwide. When using the SERVO-i on battery power, in a very small number of cases, some battery modules distributed aft
| Recall number | Z-1177-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Maquet Cardiovascular Us Sales, Llc, Wayne, NJ, United States |
| Recall initiated | 2013-03-07 |
| Classified | 2013-04-26 |
| Reported by FDA | 2013-05-08 |
| Terminated | 2016-08-11 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, AW, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JM, JP, KE, KR, KW, LK, LT, LV, MA, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SV, TH, TN, TR, TT, TW, UA, UY, VE, VN, ZA |
| Quantity | 90,000 units |
| Model numbers | 6487180 |
Product
Maquet Getinge Group battery modules for use with Maquet SERVO-i ventilator systems (Part #6487180). Battery for use with the SERVO-i Ventilator System. The SERVO-i Ventilator System is intended for treating and monitoring patients ranging from neonates to adults with respiratory failure or insufficiency. The ventilator is equipped with at least 2, NiMH battery modules (Part No. 6487180) which automatically supply 12V DC Power in case of an AC power failure, ensuring that ventilator settings and stored data remain intact in the event of an AC power failure. The battery modules give the user a possibility to use the SERVO-i during intra-hospital transportation.
Reason for recall
When using the SERVO-i on battery power, in a very small number of cases, some battery modules distributed after January 31, 2010 have a shorter battery run time than expected.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1177-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.