Data Foundry

Medical device recalls 2024

Rebstock Holding rod for fixation -For holding and fixation of… — Class II medical device recall Z-1184-2024

Ongoing recall by Medagent Inc, reported 2024-03-06, distributed in TX. Holding arm could move (twist) uncontrollably in the front area. This could lead to a serious deterioration of

Recall numberZ-1184-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedagent Inc, Portsmouth, NH, United States
Recall initiated2023-12-21
Classified2024-02-23
Reported by FDA2024-03-06
DistributedTX
Countries outside the USCN, DE, DO, DZ, GB, IN, JP, PH
Quantity84 units
Lot numbers6221672, 6221673, 6222079, 6222559, 6230282, 6230985, 6230986, 6231808, 6231989

Product

Rebstock Holding rod for fixation -For holding and fixation of flexible Leyla Arms and spatula Art. No. 06-06-285 and 06-06-285MOD. surgical instrument.

Reason for recall

Holding arm could move (twist) uncontrollably in the front area. This could lead to a serious deterioration of patient's safety if the movement is transferred to the spatula, which is attached to the holding arm.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1184-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.