Conductivity Standard Solution Catalog/Model Numbers: 10001 — Class II medical device recall Z-1185-2026
Ongoing recall by NIPRO Technical Services, Inc., reported 2026-02-04, distributed nationwide. due to a defect associated with the conductivity values of the solution used to calibrate meters and result in
| Recall number | Z-1185-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | NIPRO Technical Services, Inc., Mesa, AZ, United States |
| Recall initiated | 2025-12-12 |
| Classified | 2026-01-27 |
| Reported by FDA | 2026-02-04 |
| Distributed | Nationwide (US) |
| Quantity | 499 units |
| Model numbers | 10001, 10011, 10021, 10032, 10040, 10040C, 10041 |
Product
Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard Solutions are a secondary standard solution used for the calibration of conductivity cells together with conductivity meters
Reason for recall
due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.