Data Foundry

Medical device recalls 2013

GE — Class II medical device recall Z-1188-2013

Terminated recall by GE Healthcare, LLC, reported 2013-05-08, distributed nationwide. Fastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from

Recall numberZ-1188-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2011-07-25
Classified2013-04-29
Reported by FDA2013-05-08
Terminated2014-02-04
DistributedNationwide (US)
Countries outside the USAR, AU, BE, BR, CA, CH, CN, CO, CR, CZ, DK, EC, EG, ES, FI, FR, GB, GR, HK, HU, IE, IN, IT, JP, KR, MA, MX, NG, NO, PL, SE, SG, TR, UA
Quantity166
Reason classesdevice malfunction
Serial numbers00000220652HM2, 00000366477CN8, 00000414273CN3, 00000416117CN0, 00000416195CN6, 00000416200CN4, 00000416201CN2, 00000416202CN0, 00000416220CN2, 00000416231CN9, 00000416267CN3, 00000416268CN1, 00000416269CN9, 00000416284CN8, 00000416286CN3, 00000416295CN4, 00000416301CN0, 00000416307CN7, 00000416332CN5, 00000416333CN3, 00000416353CN1, 00000416363CN0, 00000416371CN3, 00000416372CN1, 00000416388CN7 and 141 more

Product

GE, VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspeed VCT, Discovery CT750HD. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission.

Reason for recall

Fastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from the gantry and hit the inside of the covers and other components of the gantry possibly resulting in the oil conduit breaking, spilling oil within the gantry and outside the system. This issue affects GE Healthcare VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspe