Data Foundry

Medical device recalls 2017

BD Vacutainer Buffered Sodium Citrate: — Class II medical device recall Z-1189-2017

Terminated recall by Becton Dickinson & Company, reported 2017-02-22, distributed nationwide. A portion of this lot of blood collection tubes was manufactured with approximately twice the expected amount

Recall numberZ-1189-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBecton Dickinson & Company, Franklin Lakes, NJ, United States
Recall initiated2017-01-16
Classified2017-02-10
Reported by FDA2017-02-22
Terminated2018-02-07
DistributedNationwide (US)
Countries outside the USCA, ID, MX
Quantity1,488,800 tubes
Lot numbers6090572
Model numbers369714

Product

BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (4.5 ml x 13 x 75 mm) Catalog Number: 369714 Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations

Reason for recall

A portion of this lot of blood collection tubes was manufactured with approximately twice the expected amount of buffered sodium citrate additive, an anticoagulant.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1189-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.