Offset Reamer Handle — Class II medical device recall Z-1190-2017

Terminated recall by Greatbatch Medical, reported 2017-02-22, distributed nationwide. On June 14, 2016, Greatbatch discovered a significant increase in the number of field complaints for the T5766

Recall numberZ-1190-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGreatbatch Medical, Plymouth, MN, United States
Recall initiated2016-10-19
Classified2017-02-10
Reported by FDA2017-02-22
Terminated2019-03-08
DistributedNationwide (US)
Countries outside the USAT, CA, DE, JP
Quantity1,441

Product

Offset Reamer Handle, T5766 Product Usage: The reamer handles are re-usable surgical instruments used in arthroplasty surgery of the hip joints, where either total joint replacement or resurfacing surgery is performed using a reamer cutting instrument. The reamer handles are intended for use with a surgical driver (active device). The reamer handle is connected at the distal end to a reamer. The T5766 reamer handle is an offset design with a drive shaft mechanism that spins on roller bearings and is encapsulated in an inner clam shell housing that is constrained within a Radel¿ tube and outer housing.

Reason for recall

On June 14, 2016, Greatbatch discovered a significant increase in the number of field complaints for the T5766 Offset Reamer Handle in which the U-joints of the drive chain component seizes during orthopedic procedures rendering the device non-functional. The 2016 failure rate for this failure exceeds the occurrence rate as defined in the DFMEA.