Data Foundry

Medical device recalls 2025

Brand Name: RayStation- Intended as a Radiation Therapy Treatment… — Class II medical device recall Z-1190-2025

Ongoing recall by RAYSEARCH LABORATORIES AB, reported 2025-02-26, distributed nationwide. Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evalua

Recall numberZ-1190-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmRAYSEARCH LABORATORIES AB, Stockholm, N/A, Sweden
Recall initiated2025-01-16
Classified2025-02-19
Reported by FDA2025-02-26
DistributedNationwide (US)
Countries outside the USAT, AU, BE, CA, CN, DE, DK, ES, FR, GB, HK, IN, IT, JP, KR, NL, NO, PL, SG, TH, TW
Quantity5
UPC / GTIN / UDI-DI07350002010426, 07350002010495, 07350002010501, 07350002010570, 07350002010600
Lot numbers0735000201042620211208, 0735000201049520220312, 0735000201050120220422, 0735000201057020221222, 0735000201060020220620
Serial numbers12.0.0.932, 12.0.3.68, 12.0.4.12, 12.1.0.1221, 12.3.0.119
Model numbers12.0.0.932, 12.0.3.68, 12.0.4.12, 12.1.0.1221, 12.3.0.119
Expiration dates2028-03-20

Product

Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12 and 12.3.0.119 Software Version: RayStation 11B and 11B Service Pack 1, 2, 3 and Service Pack Toshiba 1

Reason for recall

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.