Data Foundry

Medical device recalls 2023

Alaris PC Unit 8015 utilizing 802 — Class II medical device recall Z-1192-2023

Ongoing recall by CareFusion 303, Inc., reported 2023-03-08, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failu

Recall numberZ-1192-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCareFusion 303, Inc., San Diego, CA, United States
Recall initiated2023-01-24
Classified2023-03-01
Reported by FDA2023-03-08
DistributedAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY
Countries outside the USAE, AU, CA, IL, KW, NZ, QA, SA, ZA
Quantity83,185
Reason classessoftware
UPC / GTIN / UDI-DI10885403801532, 10885403801549

Product

Alaris PC Unit 8015 utilizing 802.11 b/g or 802.11 a/b/g wireless network cards, with IP addresses from x.x.x.224 to x.x.x.255

Reason for recall

Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems. Users with interfaces to electronic medication record or admin discharge transfer, may be unable to receive/send data; infusions may need to be manually programmed

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1192-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.