Data Foundry

Medical device recalls 2016

Algostim LLC — Class II medical device recall Z-1194-2016

Terminated recall by QiG Group LLC, reported 2016-03-30. The implantable pulse generator (IPG) model 2408 had suboptimal performance establishing a connection to the e

Recall numberZ-1194-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmQiG Group LLC, Blaine, MN, United States
Recall initiated2016-01-26
Classified2016-03-18
Reported by FDA2016-03-30
Terminated2017-04-05
Countries outside the USDE, LU
Quantity31
Serial numbers01DC94, 01DC98, 01DC9B, 01DCA6, 01DCB, 01DCB2, 01DCB3, 01DCB4, 01DCB5, 01DCB6
Model numbers3400

Product

Algostim LLC, Algovita External Pulse Generator, Trial Stimulator, Model 3400

Reason for recall

The implantable pulse generator (IPG) model 2408 had suboptimal performance establishing a connection to the external devices unless they were brought into close proximity (as close as 5 cm). The following functions may not be available: Reprogramming the stimulator, Adjusting stimulation parameters (on/off, intensity modifications, stimulation program selection), and Recharging the stimulator.