Abbott Alinity c Processing Module LN 03R67-01 — Class II medical device recall Z-1197-2019
Terminated recall by Abbott Gmbh & Co. KG, reported 2019-04-24, distributed nationwide. The safety interlock covering the septum piercing probes within the bulk solution bottle holder may not deploy
| Recall number | Z-1197-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abbott Gmbh & Co. KG, Wiesbaden, N/A, Germany |
| Recall initiated | 2019-02-26 |
| Classified | 2019-04-18 |
| Reported by FDA | 2019-04-24 |
| Terminated | 2022-09-28 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BD, BE, BG, BH, BR, CA, CH, CL, CN, CO, CZ, DE, DK, EE, ES, FI, FR, GB, GR, GT, HK, HR, ID, IE, IL, IN, IT, JO, JP, KE, KR, LB, LV, MX, MY, NL, NO, NZ, PH, PK, PL, PT, RO, RS, SA, SE, SG, SI, TH, TR, TT, TW, VN, ZA |
| Quantity | 559 |
Product
Abbott Alinity c Processing Module LN 03R67-01 - Product Usage: The Alinity c processing module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and potentiometric detection technology. The Alinity c processing module uses photometric detection technology to measure sample absorbance for the quantification of analyte concentration and uses potentiometric detection technology to measure the electrical potential in a sample. In addition, the Alinity c processing module uses an integrated chip technology (ICT) module to measure potentiometric assays (electrolytes).
Reason for recall
The safety interlock covering the septum piercing probes within the bulk solution bottle holder may not deploy when a bulk solution bottle is removed from the Alinity ci series and Alinity s System.