Data Foundry

Medical device recalls 2025

Hot Dog PATIENT WARMING — Class II medical device recall Z-1197-2025

Ongoing recall by Augustine Temperature Management, LLC, reported 2025-02-26, distributed nationwide. There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It tur

Recall numberZ-1197-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAugustine Temperature Management, LLC, Minnetonka, MN, United States
Recall initiated2024-12-16
Classified2025-02-20
Reported by FDA2025-02-26
DistributedNationwide (US)
Quantity4 units
UPC / GTIN / UDI-DI00855913001817
Serial numbers1304292, 1304293, 1304296, 1304297
Model numbersUS501

Product

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia

Reason for recall

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1197-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.