Data Foundry

Medical device recalls 2025

Hot Dog PATIENT WARMING — Class II medical device recall Z-1198-2025

Ongoing recall by Augustine Temperature Management, LLC, reported 2025-02-26, distributed nationwide. There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It tur

Recall numberZ-1198-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAugustine Temperature Management, LLC, Minnetonka, MN, United States
Recall initiated2024-12-16
Classified2025-02-20
Reported by FDA2025-02-26
DistributedNationwide (US)
Quantity1 units
UPC / GTIN / UDI-DI00855913001824
Serial numbers1304283
Model numbersUS502

Product

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia

Reason for recall

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.