Data Foundry

Medical device recalls 2026

Medline medical procedure convenience kits labeled as: 1) C-SECTION… — Class II medical device recall Z-1199-2026

Ongoing recall by Medline Industries, LP, reported 2026-02-04, distributed nationwide. Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to c

Recall numberZ-1199-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedline Industries, LP, Northfield, IL, United States
Recall initiated2025-12-12
Classified2026-01-28
Reported by FDA2026-02-04
DistributedNationwide (US)
Quantity96 units
Reason classespackaging
UPC / GTIN / UDI-DI10198459089206, 10198459117879, 10198459603709, 40198459089207, 40198459117870, 40198459603700
Lot numbers25KMD145, 25KMD420, 25KMI679

Product

Medline medical procedure convenience kits labeled as: 1) C-SECTION PACK, Kit SKU DYNJ35494C; 2) SMJ C-SECTION PACK-LF, Kit SKU DYNJ38582O; 3) RR-L&D C-SECTION PACK-LF, Kit SKU DYNJ82084F.

Reason for recall

Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.