Data Foundry

Medical device recalls 2017

Hercules 360 Universal Stabilizing Arm — Class II medical device recall Z-1201-2017

Terminated recall by Terumo Cardiovascular Systems Corporation, reported 2017-02-22. Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing

Recall numberZ-1201-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmTerumo Cardiovascular Systems Corporation, Elkton, MD, United States
Recall initiated2014-08-20
Classified2017-02-14
Reported by FDA2017-02-22
Terminated2017-02-15
Quantity260

Product

Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel

Reason for recall

Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device may come loose, causing the pins and sleeve to separate from the device. While the arm can still function without the pins and sleeve in place, there is potential for the loose components to fall into the surgical site.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1201-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.