The Schiller PB 1000 — Class II medical device recall Z-1203-2017

Terminated recall by Merge Healthcare, Inc., reported 2017-02-22, distributed nationwide. Non-invasive blood pressure pumps (NIBP) voltage is not sufficient and causes the NIBP pumps to fail prior to

Recall numberZ-1203-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMerge Healthcare, Inc., Hartland, WI, United States
Recall initiated2015-05-31
Classified2017-02-15
Reported by FDA2017-02-22
Terminated2020-03-13
DistributedNationwide (US)
Quantity341 sites
Serial numbers10.03.XXXXX

Product

The Schiller PB 1000, also known as: Merge Hemo Patient Data Module (PDM), V2 model, a major component of the Merge Hemo software.

Reason for recall

Non-invasive blood pressure pumps (NIBP) voltage is not sufficient and causes the NIBP pumps to fail prior to reaching adequate initial pressure.