CentraLink¿ Data Management System Software Versions: v13x — Class II medical device recall Z-1204-2017

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2017-02-22, distributed nationwide. There is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifyi

Recall numberZ-1204-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Tarrytown, NY, United States
Recall initiated2016-11-29
Classified2017-02-15
Reported by FDA2017-02-22
Terminated2018-06-07
DistributedNationwide (US)
Countries outside the USCA
Quantity3,893 units

Product

CentraLink¿ Data Management System Software Versions: v13x,v14x,v15x The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location. Siemens Healthcare Diagnostics has determined that there is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifying the sample type.

Reason for recall

There is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifying the sample type. This can occur when an order is received from the LIS without a sample type, requiring that the sample type be set in CentraLink based on the sample type of the test in the order.