Dexcom G6 — Class II medical device recall Z-1204-2025
Ongoing recall by Dexcom, Inc., reported 2025-02-26, distributed nationwide. Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/aler
| Recall number | Z-1204-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Dexcom, Inc., San Diego, CA, United States |
| Recall initiated | 2025-01-27 |
| Classified | 2025-02-20 |
| Reported by FDA | 2025-02-26 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, BE, CA, CH, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IT, JP, KR, KW, NL, NO, NZ, OM, PL, QA, SA, SE, SG, SI, SK, TR, ZA |
| UPC / GTIN / UDI-DI | 00386270001900 |
| Model numbers | STK-OE-001 |
Product
Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OE-001
Reason for recall
Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net error occurs when an alarm/alert should be triggered, the initial alarm/alert will not be delivered until a subsequent alarm/alert is triggered. At that time, the initial alarm/alert is delivered, but not the second one. Alarm/alerts will continue to be delayed and be one alarm/alert behind, causing delayed alarm/alerts until the receiver is reset. This condition could result in the missed detection of a hyperglycemic or hypoglycemic event. Second, a single EGV reading may be delayed for 5 minutes after the initial .net operating system error. If the user receives a single EGV reading delayed by 5 minutes, the delay is not expected to cause user harm.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1204-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.