Data Foundry

Medical device recalls 2014

Reusable Clip Applier Manufactured by AtriCure — Class II medical device recall Z-1209-2014

Terminated recall by Atricure Inc, reported 2014-03-26, distributed nationwide. The firm was notified by a customer of a cracked spring in the handle of the RCD1 device.

Recall numberZ-1209-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAtricure Inc, West Chester, OH, United States
Recall initiated2014-01-23
Classified2014-03-14
Reported by FDA2014-03-26
Terminated2014-05-13
DistributedNationwide (US)
Quantity46 units
Reason classespackaging
Lot numbersMFR-0424-1

Product

Reusable Clip Applier Manufactured by AtriCure, Inc., 6217 Centre Park Drive, West Chester, OH 45069. The Reusable Clip Applier (Applier) is intended to be used to deploy the Gillinov-Cosgrove LAA Clip (Clip). The Gillinov-Cosgrove LAA Clip (Clip) is an implantable device used for occlusion of the left atrial appendage.

Reason for recall

The firm was notified by a customer of a cracked spring in the handle of the RCD1 device.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1209-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.