Reusable Clip Applier Manufactured by AtriCure — Class II medical device recall Z-1209-2014
Terminated recall by Atricure Inc, reported 2014-03-26, distributed nationwide. The firm was notified by a customer of a cracked spring in the handle of the RCD1 device.
| Recall number | Z-1209-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Atricure Inc, West Chester, OH, United States |
| Recall initiated | 2014-01-23 |
| Classified | 2014-03-14 |
| Reported by FDA | 2014-03-26 |
| Terminated | 2014-05-13 |
| Distributed | Nationwide (US) |
| Quantity | 46 units |
| Reason classes | packaging |
| Lot numbers | MFR-0424-1 |
Product
Reusable Clip Applier Manufactured by AtriCure, Inc., 6217 Centre Park Drive, West Chester, OH 45069. The Reusable Clip Applier (Applier) is intended to be used to deploy the Gillinov-Cosgrove LAA Clip (Clip). The Gillinov-Cosgrove LAA Clip (Clip) is an implantable device used for occlusion of the left atrial appendage.
Reason for recall
The firm was notified by a customer of a cracked spring in the handle of the RCD1 device.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1209-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.