The Enzyme 2 Calibrator is an in vitro diagnostic product for the… — Class III medical device recall Z-1216-2017

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2017-02-22, distributed nationwide. Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vi

Recall numberZ-1216-2017
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Newark, DE, United States
Recall initiated2016-05-05
Classified2017-02-16
Reported by FDA2017-02-22
Terminated2018-02-02
DistributedNationwide (US)
Quantity13,147 units
Reason classespotency
Lot numbers4MD048, 5DD026, 5ED071, 5GD004, 5JD058, 5LD061

Product

The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.

Reason for recall

Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.