Data Foundry

Medical device recalls 2018

PERSONA¿ THE PERSONALIZED KNEE SYSTEM "PSN TIB STM 5 DEG SZ D R"… — Class II medical device recall Z-1224-2018

Terminated recall by Zimmer Biomet, Inc., reported 2018-04-04. A complaint was received for Item 42-5320-067-02, lot 63236047. It was report that there was no set screw or t

Recall numberZ-1224-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmZimmer Biomet, Inc., Warsaw, IN, United States
Recall initiated2017-12-12
Classified2018-03-27
Reported by FDA2018-04-04
Terminated2019-04-09
Quantity3 units
Lot numbers63236047
Model numbers42-5320-067-02

Product

PERSONA¿ THE PERSONALIZED KNEE SYSTEM "PSN TIB STM 5 DEG SZ D R" indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.

Reason for recall

A complaint was received for Item 42-5320-067-02, lot 63236047. It was report that there was no set screw or taper plug included in the box. Complaint investigation confirmed that the entire lot of 63236047 was missing the set screw and taper plug.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1224-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.