Data Foundry

Medical device recalls 2014

VITROS Chemistry Products White Reference Slides — Class II medical device recall Z-1226-2014

Terminated recall by Ortho-Clinical Diagnostics, reported 2014-03-26, distributed nationwide. An incorrect assay value (0.3785) was printed on the product carton for the 460 nm wavelength (filter) that ma

Recall numberZ-1226-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2014-01-09
Classified2014-03-19
Reported by FDA2014-03-26
Terminated2015-07-07
DistributedNationwide (US)
Countries outside the USAU, BR, CA, CL, CN, CO, DE, ES, FR, GB, IN, IT, JP, MX, PA, SG, VE
Quantity64 units
Reason classespotency
Lot numbers9052-0045-6358
Model numbersJ02315
Expiration dates2015-08-01

Product

VITROS Chemistry Products White Reference Slides, White Correction Factor Slide (WCF), DT Slides, 25 Slides, Part Number J02315; Made in USA by Ortho- Clinical Diagnostics, Inc. 100 Indigo Creek Drive, Rochester, NY 14626 --- This product is a part/ tool used when performing the correction factors adjustment on the VITROS Chemistry Systems. It does not have a 510(k) number. Because the WCF slide is a part/ tool, and is not a device used for diagnostic testing, there is no Intended Use statement. The following summary is provided as a description/ explanation of the WCF slide. The WCF slide is a part/ tool used in the reflectometer correction factor adjustment procedure in order to optimize the optical calibration of the VITROS 250/350/5,1 FS/4600 Chemistry Systems and the VITROS 5600 Integrated

Reason for recall

An incorrect assay value (0.3785) was printed on the product carton for the 460 nm wavelength (filter) that may cause a delay in the ability to obtain and report TBIL, Bu, Bc, and derived tests results. The correct assay value is 0.8658.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1226-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.