Back up batteries in Ventilators labeled as the following: a — Class I medical device recall Z-1226-2022

Ongoing recall by GE Healthcare, LLC, reported 2022-06-15, distributed nationwide. Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC m

Recall numberZ-1226-2022
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2022-04-18
Classified2022-06-06
Reported by FDA2022-06-15
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BD, BE, BG, BR, CA, CL, CN, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, HU, ID, IE, IN, IT, JP, KE, KR, KW, LV, MX, MY, NG, NI, NL, NO, NZ, PA, PH, PL, PT, RO, RU, SA, SE, SG, TH, TR, UY
Quantity34,940 ventilators
UPC / GTIN / UDI-DI00840682102346
Serial numbers0221, CACK00526, CBC101027, CBC101142, CBC101144, CBC101146, CBC101174, CBCC00160, CBCC00192, CBCH00170, CBCJ00102, CBCJ00106, CBCJ00107, CBCJ00108, CBCJ00110, CBCJ00113, CBCJ00145, CBCJ00149, CBCJ00150, CBCJ00151, CBCJ00160, CBCJ00162, CBCJ00174, CBCJ00176, CBCJ00178 and 475 more
Model numbers1505-9000-000, 1505-9003-000, 1506-8600-000, 1506-8600-000-XXXXXX

Product

Back up batteries in Ventilators labeled as the following: a. CARESCAPE R860 Ventilators (K142679) b. ENGSTROM (K111116) c. ENGSTROM Carestation (K111116) d. ENGSTROM PRO (K111116)

Reason for recall

Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.