Back up batteries in Ventilators labeled as the following: a — Class I medical device recall Z-1226-2022
Ongoing recall by GE Healthcare, LLC, reported 2022-06-15, distributed nationwide. Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC m
| Recall number | Z-1226-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | GE Healthcare, LLC, Waukesha, WI, United States |
| Recall initiated | 2022-04-18 |
| Classified | 2022-06-06 |
| Reported by FDA | 2022-06-15 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, BD, BE, BG, BR, CA, CL, CN, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, HU, ID, IE, IN, IT, JP, KE, KR, KW, LV, MX, MY, NG, NI, NL, NO, NZ, PA, PH, PL, PT, RO, RU, SA, SE, SG, TH, TR, UY |
| Quantity | 34,940 ventilators |
| UPC / GTIN / UDI-DI | 00840682102346 |
| Serial numbers | 0221, CACK00526, CBC101027, CBC101142, CBC101144, CBC101146, CBC101174, CBCC00160, CBCC00192, CBCH00170, CBCJ00102, CBCJ00106, CBCJ00107, CBCJ00108, CBCJ00110, CBCJ00113, CBCJ00145, CBCJ00149, CBCJ00150, CBCJ00151, CBCJ00160, CBCJ00162, CBCJ00174, CBCJ00176, CBCJ00178 and 475 more |
| Model numbers | 1505-9000-000, 1505-9003-000, 1506-8600-000, 1506-8600-000-XXXXXX |
Product
Back up batteries in Ventilators labeled as the following: a. CARESCAPE R860 Ventilators (K142679) b. ENGSTROM (K111116) c. ENGSTROM Carestation (K111116) d. ENGSTROM PRO (K111116)
Reason for recall
Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.