Data Foundry

Medical device recalls 2014

Philips HeartStart Home and OnSite — Class II medical device recall Z-1232-2014

Terminated recall by Philips Medical Systems North America Inc., reported 2014-03-26, distributed nationwide. Philips is recalling HeartStart Home and OnSite (HS1) automated external defibrillators (AEDs) because one or

Recall numberZ-1232-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Medical Systems North America Inc., Bothell, WA, United States
Recall initiated2014-02-06
Classified2014-03-19
Reported by FDA2014-03-26
Terminated2017-03-23
DistributedNationwide (US)
Model numbersM5066A, M5067A, M5068A

Product

Philips HeartStart Home and OnSite (HS1) automated external defibrillators (AEDs). Product Usage: The models M5066A, M5067A, and M5068A are designed for use on a person in sudden cardiac arrest who is: -Unresponsive when shaken -Not breathing normally If in doubt, apply the pads. A training class in CPR/AED use is recommended for anyone who may use the HS1. If the person is an infant or child younger than eight years old or weighs less than 55 lbs (25 kg), use the special infant/child pads. If the child appears older/larger, use the adult pads. Do not delay treatment to determine the child s exact age or weight. If in doubt, use the adult pads. Models M5066A (for HS1 Onsite), M5067A (OnSite) , and M5068A (for HS1 Home).

Reason for recall

Philips is recalling HeartStart Home and OnSite (HS1) automated external defibrillators (AEDs) because one or both contact pins may be contaminated by residue from the soldering process. This contamination could prevent an adequate connection between the pads cartridge and the AED. If poor contact between the pads cartridge and AED occurs, the device may be unable to deliver therapy in an emergency.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1232-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.