syngo X Workplace is a medical workstation for real-time viewing — Class II medical device recall Z-1232-2016
Terminated recall by Siemens Medical Solutions USA, Inc, reported 2016-03-30, distributed nationwide. After importing, the segmentation results appear mirrored at the CARTO system and can't be used for the ablati
| Recall number | Z-1232-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Medical Solutions USA, Inc, Malvern, PA, United States |
| Recall initiated | 2016-02-24 |
| Classified | 2016-03-23 |
| Reported by FDA | 2016-03-30 |
| Terminated | 2017-02-16 |
| Distributed | Nationwide (US) |
| Quantity | 6 |
| Serial numbers | 12177, 20026, 20041, 20098, 20119, 20124 |
Product
syngo X Workplace is a medical workstation for real-time viewing, image manipulation, 3D-visualization, communication, and storage of medical images and data on exchange media. It is used for diagnostic image viewing and post processing during interventional procedures. .
Reason for recall
After importing, the segmentation results appear mirrored at the CARTO system and can't be used for the ablation procedure.