VITROS XT 3400 Chemistry System — Class II medical device recall Z-1232-2024
Ongoing recall by Ortho-Clinical Diagnostics, Inc., reported 2024-03-13, distributed nationwide. A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configu
| Recall number | Z-1232-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ortho-Clinical Diagnostics, Inc., Rochester, NY, United States |
| Recall initiated | 2024-01-26 |
| Classified | 2024-03-01 |
| Reported by FDA | 2024-03-13 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, BE, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PT, RU, SE, SG |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 10758750031986 |
| Serial numbers | J34500123, J34500128, J34500129, J34500130, J34500140, J34500180, J34500190, J34500257, J34500261, J34500269, J34500271, J34500289, J34500292, J34500301, J34500319, J34500337, J34500356, J34500363, J34500386, J34500422, J34500447, J34500473, J34500474, J34500510, J34500515 and 26 more |
| Model numbers | 6844458 |
Product
VITROS XT 3400 Chemistry System, Catalog No. 6844458
Reason for recall
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.