Data Foundry

Medical device recalls 2017

The Sarns Centrifugal Pump is a single use device sterilized by… — Class II medical device recall Z-1235-2017

Terminated recall by Terumo Cardiovascular Systems Corporation, reported 2017-03-01. Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (

Recall numberZ-1235-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmTerumo Cardiovascular Systems Corporation, Elkton, MD, United States
Recall initiated2015-05-27
Classified2017-02-22
Reported by FDA2017-03-01
Terminated2017-02-23
Quantity579 units
Reason classespackaging

Product

The Sarns Centrifugal Pump is a single use device sterilized by ethylene oxide gas. The device is sold single sterile in cartons of 8 units, or may be provided in bulk, non-sterile to another Terumo location for inclusion in Cardiovascular Procedure Kits. Cardiovascular Procedure kits are packaged individually.

Reason for recall

Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the bypass circuit and during CPB procedures. In each of the reported incidents, there were no reports of patient injury as a result of the centrifugal pump leaks. Terumo Cardiovascular System's investigation into the incident has identified the cause of the leaks to be cracks in both the Top Housing and Magnet Back Housing of the pump. When the cracks cause the housing to become compromised, it is possible for fluid to leak from the pump head.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1235-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.