Data Foundry

Medical device recalls 2013

Stryker Irrigator Reusable Tips: 3mm X 32cm Regular Tip — Class II medical device recall Z-1236-2013

Terminated recall by Stryker Endoscopy, reported 2013-05-15, distributed nationwide. The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed

Recall numberZ-1236-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker Endoscopy, San Jose, CA, United States
Recall initiated2013-03-27
Classified2013-05-03
Reported by FDA2013-05-15
Terminated2013-09-24
DistributedNationwide (US)
Countries outside the USAU, CA, CL, CN, FR, GR, IN, JP, KR, MX, NL, NZ, PT, RO, SG, TW, ZA
Quantity147,921
Model numbers250-070-490

Product

Stryker Irrigator Reusable Tips: 3mm X 32cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.

Reason for recall

The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.