SIGNA Creator — Class II medical device recall Z-1239-2025

Ongoing recall by GE Healthcare (China) Co., Ltd., reported 2025-03-05, distributed nationwide. For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" i

Recall numberZ-1239-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare (China) Co., Ltd., Beijing, N/A, China
Recall initiated2025-01-31
Classified2025-02-27
Reported by FDA2025-03-05
DistributedNationwide (US)
Quantity2,538
UPC / GTIN / UDI-DI00082427010138, 00082427010169, 00082427010183, 00082427020090, 00082427020106, 00082427030198, 00082427040203, 00082427040272, 00082427040340, 00082427060102, 00082427060119, 00082427080100, 00082427080131, 00082427090420, 00082427090444, 00082427090505, 00082427100334, 00082427100396, 00082427100440, 00082427100457, 00082427100471, 00082427110159, 00082427120226, 00082427120288, 00082427120325 and 43 more

Product

SIGNA Creator, SIGNA Explorer, NMRI systems

Reason for recall

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.