Data Foundry

Medical device recalls 2014

2 — Class II medical device recall Z-1247-2014

Terminated recall by Stryker Instruments Div. of Stryker Corporation, reported 2014-03-26, distributed in CA, CO, ID, IN, NV, OR, SD, TN…. Stryker Instruments is voluntarily recalling the single use 2.5MM Aggressive Blade because the line on the aff

Recall numberZ-1247-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker Instruments Div. of Stryker Corporation, Portage, MI, United States
Recall initiated2014-03-18
Classified2014-03-20
Reported by FDA2014-03-26
Terminated2014-05-15
DistributedCA, CO, ID, IN, NV, OR, SD, TN, WA
Countries outside the USMY
Quantity90 units
Lot numbers13280CG2

Product

2.5MM Aggressive Blade Known as a Sinuscopic Shaver Blade, this device is composed of a metal cutter and plastic drive shaft that goes into a metal housing supported by a plastic hub. ENT Precision Cutter, Aggressive, ESSxlHUMMER 4, Rx Only It is used for ENT procedures and functions by means of rotation when adapted to a handpiece and console. Through the rotation, the necessary tissue is resected as required by the user. The blade housing has an etching line guide that starts with a straight line and along with a scale from 1cm to 7cm in 1cm increments that serves the doctor as a guide by indicating how far the device is being introduced into the surgical area.

Reason for recall

Stryker Instruments is voluntarily recalling the single use 2.5MM Aggressive Blade because the line on the affected part (P/N 5290-928-000, Lot #13280CG2) was a straight line without a number scale or logo. The line required for this part number includes a number scale that goes from 1cm  7cm.