Data Foundry

Medical device recalls 2015

RAPIDLab 1245 System — Class II medical device recall Z-1254-2015

Terminated recall by Siemens Healthcare Diagnostics Inc, reported 2015-03-18, distributed nationwide. Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater th

Recall numberZ-1254-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics Inc, Norwood, MA, United States
Recall initiated2014-12-30
Classified2015-03-06
Reported by FDA2015-03-18
Terminated2017-08-02
DistributedNationwide (US)
Countries outside the USAE, AR, AT, BE, BG, BH, BR, CA, CH, CL, CO, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GR, HR, HU, IE, IL, IQ, IS, IT, KW, LB, LT, LU, LV, MX, NL, NO, PE, PL, PT, QA, RO, RS, RU, SA, SE, SI, SK, TR, VE, ZA
Quantity241 devices

Product

RAPIDLab 1245 System; Siemens Material Number: 10321844, 10491393, 10337179. The Rapidlab 1200 systems are intended for in vitro diagnostic use by healthcare professionals in the quantitative testing of human whole blood. The systems can determine the following parameters: RAPIDLab 1245 pH, pC02, p02, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb RAPIDLab 1265 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb.

Reason for recall

Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1254-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.