Data Foundry

Medical device recalls 2017

Siemens ADVIA Centaur Systems Calibrator A Kits — Class II medical device recall Z-1255-2017

Terminated recall by Siemens Healthcare Diagnostics, Inc, reported 2017-03-08, distributed nationwide. Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A

Recall numberZ-1255-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Recalling firmSiemens Healthcare Diagnostics, Inc, East Walpole, MA, United States
Recall initiated2017-01-10
Classified2017-02-25
Reported by FDA2017-03-08
Terminated2019-07-16
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, BD, BE, BG, BH, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DO, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HR, HU, ID, IE, IL, IN, IQ, IT, JO, JP, KR, LK, LT, LU, LV, MT, MX, MY, NI, NL, NO, NP, OM, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, TH, TN, TR, TW, UY, VN, ZA
Quantity1,937 kits
Reason classespotency
Lot numbers07996A90, 2017/07/24, 21172A90, 27221A90, 44366A90, 63785A90
Model numbers04800646, 04800735, 04800840, 10285903, 10285904, 10285905

Product

Siemens ADVIA Centaur Systems Calibrator A Kits, Cat No. 04800646, SMN 10285903 (2 pack)Cat No. 04800735, SMN 10285904 (6 pack) Cat No. 04800840, SMN 10285905 (6 pack - REF)

Reason for recall

Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator RLU %CVs when using Calibrator A kit lots ending in 90 with the FT4 assay. The performance of the FT4 assay when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur CP System is not affected. Customers may continue to use Calibrator A kit lots ending in 90 for the ADVIA Centaur Systems FT3, T3, T4 and TUp assays.