Data Foundry

Medical device recalls 2018

Altivate Anatomic — Class II medical device recall Z-1259-2018

Completed recall by Encore Medical, Lp, reported 2018-04-04, distributed nationwide. It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Gleno

Recall numberZ-1259-2018
Product typeMedical device
ClassificationClass II
StatusCompleted
Voluntary or mandatedVoluntary
Recalling firmEncore Medical, Lp, Austin, TX, United States
Recall initiated2018-01-29
Classified2018-03-28
Reported by FDA2018-04-04
DistributedNationwide (US)
Quantity92 units
Model numbers804-07-500

Product

Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.

Reason for recall

It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1259-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.