Altivate Anatomic — Class II medical device recall Z-1260-2018
Completed recall by Encore Medical, Lp, reported 2018-04-04, distributed nationwide. It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Gleno
| Recall number | Z-1260-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Completed |
| Voluntary or mandated | Voluntary |
| Recalling firm | Encore Medical, Lp, Austin, TX, United States |
| Recall initiated | 2018-01-29 |
| Classified | 2018-03-28 |
| Reported by FDA | 2018-04-04 |
| Distributed | Nationwide (US) |
| Quantity | 93 units |
| Model numbers | 804-07-540 |
Product
Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 54mm, REF 804-07-540 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
Reason for recall
It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1260-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.