Data Foundry

Medical device recalls 2019

Discovery NM/CT 670 ES Product Usage: The GE Discovery NM/CT 670… — Class II medical device recall Z-1260-2019

Terminated recall by GE Healthcare, LLC, reported 2019-05-08. During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector scr

Recall numberZ-1260-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2019-03-20
Classified2019-05-02
Reported by FDA2019-05-08
Terminated2020-10-05
Countries outside the USAT, CA, CH, DE, ES, FR, GB, HR, IN, IT, JP, KR, NL, PL, PT, SE, SK, ZA
Quantity3 units
Model numbers403777D670C, A51773134, DE203284, ESDX35020, ESPX54001, ESPY54002, H3100BL, H3100BM

Product

Discovery NM/CT 670 ES Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce:

Reason for recall

During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1260-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.