Data Foundry

Medical device recalls 2018

Syngo — Class II medical device recall Z-1261-2018

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2018-04-04. When the archiving configuration is changed, data received/created after the upgrade may be flagged as "Not to

Recall numberZ-1261-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2018-01-10
Classified2018-03-28
Reported by FDA2018-04-04
Terminated2018-10-02
Quantity165 units
Reason classessoftware
Model numbers10496180

Product

Syngo.via software is intended to be used for viewing, manipulation, communication, and storage of medical images. It can be used as a standalone device or together with a variety of cleared and unmodified syngo based software options.

Reason for recall

When the archiving configuration is changed, data received/created after the upgrade may be flagged as "Not to be archived". This is caused by the automatic function for cleaning up temporary data being disabled by the software upgrade. Due to the disabled cleanup function, disc capacity for free space decreases faster than usual. Unless the archiving configuration and the cleanup automatic function are reverted back to the original state (prior to the software upgrade), data that was incorrectly flagged "Not to be archived" must be manually prevented from being deleted by either clinical administrator or service engineer.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1261-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.