Data Foundry

Medical device recalls 2019

NM 830 Product Usage: The Discovery NM 630 is an all-purpose dual… — Class II medical device recall Z-1261-2019

Terminated recall by GE Healthcare, LLC, reported 2019-05-08. During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector scr

Recall numberZ-1261-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2019-03-20
Classified2019-05-02
Reported by FDA2019-05-08
Terminated2020-10-05
Countries outside the USAT, CA, CH, DE, ES, FR, GB, HR, IN, IT, JP, KR, NL, PL, PT, SE, SK, ZA
Quantity37 units
Model numbers0850260165, 0850260166, 100039NU05, 225761NM, 225761NM1, 256737NM, 304526D830, 361275NM830, 362759NU01, 410554NM, 479338N830, 5376204-32-1, 573458NM, 585396NM, 585786NM830, 603742NM830, 615316NM830, 631726D830, 705743DNM830, 716862NM830, 718238D830, 752030NU01, 772468NM830, 813315D830, 813874D830 and 51 more

Product

NM 830 Product Usage: The Discovery NM 630 is an all-purpose dual detector nuclear imaging system intended for general nuclear medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The main components of the Discovery NM 630 system are: NM Gantry with Dual detector heads, patient table, Remote Control Unit and NM operation console.

Reason for recall

During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1261-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.