VITROS Hand-held Barcode Scanner — Class II medical device recall Z-1263-2015
Terminated recall by Ortho-Clinical Diagnostics, reported 2015-03-18, distributed nationwide. In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanner
| Recall number | Z-1263-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ortho-Clinical Diagnostics, Rochester, NY, United States |
| Recall initiated | 2014-12-29 |
| Classified | 2015-03-11 |
| Reported by FDA | 2015-03-18 |
| Terminated | 2017-03-27 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, GB |
| Lot numbers | 0914F11968, 0914F11969, 0914F11970, 0914F11971, 0914F11972, 0914F11973, 0914F11974, 0914F11975, 0914F11976, 0914F11977, 0914F11978, 0914F11979, 0914F11980, 0914F11981, 0914F11982, 0914F11983, 0914F11984, 0914F11985, 0914F11986, 0914F11987, 0914F11988, 0914F11989, 0914F11990, 0914F11997, 0914F12007, 0914F12016, 0914F12017, 0914F12020, 0914F12022, 0914F12028, 0914F12033, 0914F12037, 0914F12039, 0914F12040, 0914F12041, 0914F12046, 0914F12048, 0914F12049, 0914F12050, 0914F12053 and 32 more |
| Model numbers | 6844210 |
Product
VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITROS Systems: VITROS 5,1 FS Chemistry System, VITROS 3600 Immunodiagnostics System, VITROS 4600 Chemistry System, VITROS 5600 Integrated System. This product is an accessory for use with VITROS Analyzer Systems.
Reason for recall
In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanner unexpectedly skips sample cup positions within a sample tray. This issue only occurs when using the Batch Programming option; all other sample programming options function as intended.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1263-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.