Vanguard XP Knee System Porous Femoral Component various sizes and… — Class II medical device recall Z-1265-2016
Terminated recall by Biomet, Inc., reported 2016-04-06, distributed in UT. The anterior/posterior (AP) dimension may be oversized by 0.016. If the femoral component does not fit as tig
| Recall number | Z-1265-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Biomet, Inc., Warsaw, IN, United States |
| Recall initiated | 2016-02-05 |
| Classified | 2016-03-25 |
| Reported by FDA | 2016-04-06 |
| Terminated | 2017-03-28 |
| Distributed | UT |
| Countries outside the US | NL |
| Quantity | 890 |
| Lot numbers | 195929 |
Product
Vanguard XP Knee System Porous Femoral Component various sizes and orientation a total knee replacement system. prosthesis, knee, patella/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer
Reason for recall
The anterior/posterior (AP) dimension may be oversized by 0.016. If the femoral component does not fit as tightly as intended, bony ingrowth may not occur causing loosening of the femoral component. A revision surgery may be necessary.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1265-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.