NM/CT 870 DR Product Usage: The GE Discovery NM/CT 670 system is a… — Class II medical device recall Z-1265-2019
Terminated recall by GE Healthcare, LLC, reported 2019-05-08. During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector scr
| Recall number | Z-1265-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | GE Healthcare, LLC, Waukesha, WI, United States |
| Recall initiated | 2019-03-20 |
| Classified | 2019-05-02 |
| Reported by FDA | 2019-05-08 |
| Terminated | 2020-10-05 |
| Countries outside the US | AT, CA, CH, DE, ES, FR, GB, HR, IN, IT, JP, KR, NL, PL, PT, SE, SK, ZA |
| Quantity | 27 units |
| Model numbers | 870Y64037 |
Product
NM/CT 870 DR Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce
Reason for recall
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.