Data Foundry

Medical device recalls 2023

Atellica CH Triglycerides_2: in vitro diagnostic use in the… — Class II medical device recall Z-1271-2023

Ongoing recall by Siemens Healthcare Diagnostics, Inc., reported 2023-09-06, distributed nationwide. Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglyc

Recall numberZ-1271-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Tarrytown, NY, United States
Recall initiated2023-01-27
Classified2023-08-29
Reported by FDA2023-09-06
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BR, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EG, ES, FI, FR, GB, GR, HK, HR, HU, IE, IL, IN, IT, JP, KR, KW, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, TH, TW, UA, UY, VN, ZA
Quantity5,315 units
UPC / GTIN / UDI-DI00630414610955

Product

Atellica CH Triglycerides_2: in vitro diagnostic use in the quantitative determination of triglycerides in human serum and plasma (lithium heparin, sodium heparin, potassium EDTA) using the Atellica CH Analyzer. SMN: 11537222

Reason for recall

Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1271-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.