CORTRAK* 2 Nasogastric/Nasointestinal — Class II medical device recall Z-1277-2018
Terminated recall by Halyard Health, Inc, reported 2018-04-11, distributed nationwide. The transmitting stylet within CORTRAK* 2 nasogastric/Nasointestinal (NG/NI) Feeding Tubes with ANTI-IV* conne
| Recall number | Z-1277-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Halyard Health, Inc, Alpharetta, GA, United States |
| Recall initiated | 2017-12-13 |
| Classified | 2018-03-30 |
| Reported by FDA | 2018-04-11 |
| Terminated | 2021-11-04 |
| Distributed | Nationwide (US) |
| Countries outside the US | DE |
| Lot numbers | 7839, 7876, 7909 |
Product
CORTRAK* 2 Nasogastric/Nasointestinal (NG/NI) Feeding Tubes with Electromagnetic Transmitting Stylet and CORFLO* Enteral Feeding tube, ANTI-IV* Connector, 10 Fr, 140 cm The device is used in patients that require intermittent or continuous tube feedings via the nasogastric or nasoenteric route.
Reason for recall
The transmitting stylet within CORTRAK* 2 nasogastric/Nasointestinal (NG/NI) Feeding Tubes with ANTI-IV* connector (Product Code 20-9551Trak2A) may be visible within the window of the exit port of the feeding tube. Based on specifications, the transmitting stylet in the NG tube should not be visible within the port window.