2008 K2 Hemodialysis Machine with software version 5 — Class II medical device recall Z-1278-2018

Terminated recall by Fresenius Medical Care Renal Therapies Group, LLC, reported 2018-04-11, distributed in AR, CA, CO, FL, GA, IL, MA, MD…. When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the

Recall numberZ-1278-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmFresenius Medical Care Renal Therapies Group, LLC, Waltham, MA, United States
Recall initiated2017-12-14
Classified2018-03-30
Reported by FDA2018-04-11
Terminated2021-09-22
DistributedAR, CA, CO, FL, GA, IL, MA, MD, MN, NC, NY, OH, PA, SC, TN, TX
Quantity127
UPC / GTIN / UDI-DI00840861100859, 00840861100866
Serial numbers00840861100859PLS, 0K0S151801, 190610, 1K0S169898, 1K0S169954, 1K0S170154, 1K0S170234, 1K0S171275, 1K0S171286, 1K0S171421, 1K0S171436, 2K0S176243, 2K0S176244, 2K0S176257, 2K0S176264, 2K0S176271, 2K0S176273, 2K0S176278, 2K0S176280, 2K0S176281, 2K0S176283, 2K0S176286, 2K0S176443, 2K0S176472, 2K0S176633 and 212 more

Product

2008 K2 Hemodialysis Machine with software version 5.40, Models: (1) Hemodialysis SYS OLC/Diasafe PLS 190610, UDI 00840861100859 (+Serial Number), (2) Machine Short Cap, OLC.DP HP 190630, UDI 00840861100866 (+Serial Number) ndicated for acute and chronic dialysis therapy.

Reason for recall

When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis) program the display will show an invalid message and the UF pump will run at the recirculation UF rate which may be up to 4000ml/hour. The SLED program is required to limit the UF rate to a maximum of 1000 ml/hour.